Mianserin

Mianserin Chromatogram

 

Sample: Mianserin 56 ng/ml
Inj. vol.: 500 µl serum
Extraction column : BioTrap 500 MS, 20x4.0 mm
Extraction mobile phase: 4% 2-propanol in 10mM ammonium acetate buffer, pH 10 (adjusted with ammoniumhydroxide)
Flow: 3.2 ml/min
Analytical column: Hypersil Elite C18, 150x4.6 mm, 5 µm + Elite guard, 10x4 mm, 5 µm
Analytical mobile phase: 40% acetonitrile in 50 mM formic acid
Flow: 1.0 ml/min
Detection: Post column derivatization: Beam Boost UV= 254 nm 20 m x 0.3 mm I.D. Fluorescense: ex= 270 nm, em= 430 nm

Analysis Program

Switching valve position Analysis step Time
1. Extraction position sample extraction 2.5 min
2. Elution position transfer of analyte 4 min
3. Extraction position separation and reequilibration 7.5 min

 

If you have any questions regarding the above, don´t hesitate to contact Chrom Tech!