Conductivity Detector

New! Microprocessor-based LC Detector for Ionic Species

Conductivity Detector

  • Turns any HPLC into an Ion Chromatograph
  • High Sensitivity Detection of all Charged Species
  • Temperature-Controlled Cell for Stable Baselines
  • Suitable for Suppressor-Based and Single-Column IC

 

The Best Detector for Ion Analysis

Turn your HPLC System into a High-Sensitivity Ion Chromatograph!

The Conductivity Detector provides universal, reproducible, high-sensitivity detection of all charged species. Use your HPLC system with the Conductivity Detector for simple, reliable, and inexpensive quantification of anions, cations, metals, organic acids, and surfactants down to ppb level.

Microprocessor-Based for High Sensitivity and Extended Linearity

The new advanced microprocessor-based circuit improves detection sensitivity by reducing baseline noise and drift. Linear response extends over a wide range of analyte concentrations.

Electroless Nickel Solution

Temperature-Controlled Cell for Stable, Drift-Free Baselines

Temperature fluctuations cause baseline noise and drift with other conductivity detectors. The temperature-controlled cell and heat exchanger are isolated from external temperature changes for stable baselines. Adjust or monitor cell temperature (ambient - 80º C) easily from the detector's front panel.

Baseline Stability Test

Works with any IC Method

The digital autozero in the Conductivity Detector offsets high conductivity eluants (up to 10,000 µS/cm) within 5 seconds, for detection with single-column or suppressor-based methods. Other detectors designed for suppressor-based IC systems are unable to offset the high conductivity eluants used in SCIC.

ppb level cations

 

Easy to Use, Troubleshoot, and Maintain

The Conductivity Detector makes IC analysis easier and more reliable. An intuitive membrane key-panel controls all detector functions. Three working modes (CALIBRATE, BACKGROUND, and MEASURE) let you check detector electronics, red the eluant's absolute conductivity, or perform analyses, respectively. Use the CALIBRATE mode to check the detector's calibration and gain for compliance with ISO-9000, GLP, or GMP programs. Use the BACKGROUND mode to check the quality and accuracy of the eluant. The large LED display shows detector signal, background conductance, cell temperature, or cell temperature set point. Two keys allow you to set range from 0.1 to 5000 µS/cm or cell temperature from ambient to 80º C. An RS-232 serial port provides digital communication and control via an external computer.

CONDUCTIVITY DETECTOR SPECIFICATIONS

Type: Alternating current, Synchronous Detection, 10 kHz Frequency
Range: 0.1-5000 m S/cm (DI water, 1 mL/min., 35° C)
Noise: < 0.0025 m S/cm (DI water, 1 mL/min., 35° C)
Drift: < 0.03 m S/cm/hr (DI water, 1 mL/min., 35° C)
Digital autozero: Offsets up to 10,000 m S/cm
Cell Type: 2 electrode, 0.5 m L volume, gold-plated stainless-steel electrodes
Display: 4-digit LED in 6 modes

Working Modes:
a. Calibrate: test reading of 100
m S/cm, with 10-step gain test
b. Background: absolute conductivity of eluant
c. Measure: conductivity of eluant with background offset

Display Modes:
a. Conductivity in
m S/cm
b. Temperature: actual cell temperature in
° C
c. Temperature set: cell temperature set point in
° C

Analog Signal Output: 10 mV and 1V FS, switch-selectable
Digital Communications: RS-232 output of detector signal, external set of detector functions
Time Constant: Fast or Slow, switch-selectable

CELL COMPARTMENT

Insulation: Minimum 1 inch all sides
Heater: 45W DC heating element vulcanized to aluminum block
Size: 2.7" W x 2.45" H x 6.2" L
Temperature: Factory-set at 35° C; User-adjustable ambient to 80° C
Overtemp Protection: Factory-set at 60° C; User-adjustable
Precision: ± 0.05° C
Heat Exchanger: TFE tubing knitted and encased in high thermal-mass housing

ELECTRICAL/ELECTRONIC

Power: 110/120 Vac, 50/60 Hz (field-selectable)
Data: Analog output, 0-1 Vdc or 0-10 mVdc (selectable)
Computer Interface: RS-232 serial communications

FLUID PATH

Construction: All wetted parts are inert polymers: PEEK (polyetherether ketone), and TeflonÒ

FRONT PANEL FUNCTIONS

Conductivity Detector

COND Displays conductivity in m S/cm
MEASURE Displays conductivity with background subtracted
BACKGROUND Displays absolute conductivity with no offset
CALIBRATE Checks gain and calibration of detector electronics
TEMP Displays actual cell temperature
TEMP SET Displays cell temperature set point. Adjust by using the UP and DOWN arrow keys.
AUTOZERO Digitally offsets the background conductivity while in the MEASURE mode
UP and DOWN arrows Select the range in the COND mode or the cell temperature set point while in the TEMP SET mode

CONDUCTIVITY DETECTOR

Features and Benefits

Feature Benefit
Convenience and Reliability  
Tactile membrane front panel Simple, intuitive control of all detector functions
Internal calibration Verifies electronic calibration and gain for ISO-9000, GLP, and GMP compliance
Background mode Measures eluant's absolute conductivity for quick troubleshooting
Read and set cell temperature from the front panel Fast, easy adjustment and verification of cell temperature
Remote operation of detector functions or digital transmission of detector signal via RS-232 Operate the detector of collect data with an external computer
Performance  
Microprocessor-based Low noise and drift for higher sensitivity, wide dynamic range, and fast, digital autozero
Temperature-controlled cell housing with built-in heat exchanger Stable baselines, even at high sensitivity ranges
Digital autozero Offsets high-conductivity eluants for compatibility with single-column or suppressor-based methods
Synchronous AC detection Simple two-electrode cell for high reliability and low maintenance
Versatility  
Compact modular design Adds IC capability to any existing HPLC system
Electronic offset of high-conductivity or low-conductivity eluants Works with both single-column and suppressor-based methods

Ordering information

Part No. Description
AL5500 Conductivity Detector
AL3351 Chemical suppressor module for Conductivity Detector
AL-535100 Eris 1000™ Autosuppressor

Return to Detector Home Page

Home Page Link Icon

click here to return to the home page

Home Page | Cyber Specials | What's New | Product Line | Instrumentation | Suppliers | Order Information | E-mail Us