ECD-200 Electrochemical Detector

electrochemical detector

High sensitivity and selectivity for

automated, analytical HPLC

User friendly

The ECD-200 Electrochemical Detector provides high sensitivity for automated, analytical chromatography of electroactive compounds. With 0.05 to 200 nA/V detection range, this amperometric detector offers both high sensitivity and high selectivity.

The ECD-200 Electrochemical Detector is designed for repetitive applications and increased ease of use. Two methods can be stored.

The digital display shows all important parameters for efficient interpretation:

  • volt or nA/V output data
  • volt or nA/V offset
  • potential
  • time constraint
  • base line decay
  • sensitivity
  • chromatogram

    Additional features of the ECD-200 include:

  • Optimized flow path and low dead volume of the thin layer type electrochemical cell.
  • An autozero function and easy-to-use software for automated, unattended operation.
  • LCD screen that displays the real-time chromatogram and operating parameters.
  • Control by computer or through RS-232 interface when integrated into a complete HPLC system.

Sensitive and selective
The ECD-200 provides sensitive, selective detection. The electronic transfer between electrode and compound in solution occurs only if the working potential is held at the right value to oxidize or reduce the compound; the detector is blind when the compounds react above this value.

Innovative electronics and specially designed cell
With a low dead volume, the cell maintains efficiency of the separation. The mirror polish of all cell parts eliminates hydrodynamic disturbances.

Specifications
Working Electrode Glassy carbon, flat circular, inserted in lower PCTFE half-block of the flow cell
Auxiliary Electrode Upper stainless steel half-block
Reference Electrode Ag/AgCl/Cl-polypropylene, screwed into auxiliary electrode
Spacer Seal PTFE, 0.1 mm thickness
Cell Volume 2 µl from cell inlet to working electrode
Dummy Cell Built-in, simulates electrochemical cell to test baseline noise and drift
Display LCD screen for real-time display of control parameters and chromatogram
Software Keypad entry of parameters (10 soft keys)
Working Potential -2.55 V to +2.55 V, 10 mV increments
Offset Potential -10 V to +10 V, 10 mV increments
Sensitivity Range 0.05 to 200 nA/V (12 steps)
Time Constant 0.5, 2, and 5 sec.
Manual Control Autozero and event marker
Protocol Storage Two sets of parameters can be stored
Noise Less than 3 percent short term, peak-to-peak (following ASTM standards)
Drift Less than 3 % per hour at stable temperatures
Remote Control 9-pin RS-232 connector
Electrical Power 110-220 V, 50-60 Hz
Manufacturing Standards

Safety:

EMC:

conforms to IEC 1010-1, UL 1262 and CSA 22.2-151

conforms to EN 50081-1 for emission and EN 50082-1 for immunity

Dimensions (W x D x H) 34 x 31 x 14 cm; 13.4 x 12.2 x 5.5 inches
Weight 4 kg; 8.8 lbs.
   

The operational kit includes a complete cell 2 reference electrodes, fittings, gaskets. Different working electrodes are available.

Ordering information:

Part Number Description
PI-105 ECD-200 electrochemical detector

Return to Detector Home Page

Home Page | Cyber Specials | What's New | Product Line | Instrumentation | Suppliers | Order Information | E-mail Us