Dextromethorphan

| Sample: |
Dextromethorphan (55
ng/ml) in serum |
| Inj. vol.: |
200 µl |
| Extraction column
: |
BioTrap 500 C8, 20x4.0 mm |
| Extraction mobile
phase: |
4% 2-propanol and 5 mM
sod. octanesulfonic acid in 20 m M sod. ph. b. pH 7.0 |
| Analytical column: |
Zorbax XDB-C8, 150 x 4.6 mm, 5 µm +
guard |
| Analytical mobile phase: |
30% acetonitrile and 2 mM sod.
octanesulfonic acid in 116 mM sod.ph.b. pH 2.8 |
| Detection: |
Fluorescence, ex 220 nm, 305 nm |
Analysis Program
| Switching valve position |
Analysis step |
Time |
| 1. Extraction position |
sample extraction |
3 min |
| 2. Elution position |
transfer of analyte |
5 min |
| 3. Extraction position |
separation and reequilibration |
7 min |
If you have any questions regarding the above,
don´t hesitate to contact Chrom Tech!
Ordering
information
Return to Chrom Tech´s
BioTrap 500 Homepage